Halogen-quenched uncompensated GM tube with thin mica
window 1.5-2.0 mg/cm2 thick.
Effective diameter of window is 9.1 mm (.36 in.).
Operating Range
0-.5, 0-5, 0-50 mR/hr
0-500, 0-5,000, 0-50,000 CPM
or 0-5, 0-50, 0-500 μSv/hr (SI Scale Meter Option)
Range Switch
X1, X10, X100, Battery Check.
Energy Sensitivity
1000 CPM/mR/hr (Cs137).
Detects alpha down to 2.5 MeV; typical detection efficiency at 3.6 MeV is greater than 80%. Detects beta at 50 keV with typical 35% detection efficiency and beta at 150 keV with typical 75% detection efficiency. Detects gamma and x-rays down to 10 keV typical through the window, 40 keV minimum through the case
Accuracy
Reference to Cs137
Typically ±15% from factory, ±10% with NIST Source Calibration NIST Calibration Additional
Display
Analog Meter holds full scale in fields as high as 100X maximum reading. CPM & mR/hr scale. Optional SI Scale Meter Available
Audio Indicator
Internally mounted beeper (can be turned off for silent operation).
Outputs
3.5 mm jack with pulse out
Power Requirements
9-volt alkaline battery (included). Approx. 2,000 hrs. at normal background